Summary
Type: SPE
Application #: 300600
Catalog#: 730015 - SPE Cartridges, Chromabond, C18 ec, 6ml, PP, 1000mg, 30/pk
Detection:
Conditions: Column conditioning:
2 column volumes methanol, then 1 column volume 0.05 M sodium dihydrogen
phosphate solution
Sample application:
Slowly force or aspirate 10 ml clear sample solution (2 – 5 μg vitamin B12)
through the column
Column washing:
1 column volume water; the column must not run dry!
Elution:
2 ml ethanol / 0.05 M sodium dihydrogen phosphate solution (1:1, v/v), then 3
ml dist. water
Interfering dyes can be removed using a CHROMABOND® SB (= SAX) column (3 ml /
500 mg) on top of the C18 ec column.
Authors: Macherey-Nagel